Feature:


Specification£º


Display 3 1/2 digit LCD with max. reading of 1999.
Polarity Automatic, (-) negative polarity indication.
Zero adjustment Automatic.
Over range indication Only the MSD ¡°1¡± is displayed.
Power Single, standard 9 volt battery NEDA 1604, JIS 006P, IEC6F22.
Dimension 104 (W) x 145.5 (H) x 56 (D) mm.
Weight Approx. 320gm. (including battery and packaging).
Accuracies are ¡À(% of reading + number of least significant digits) at 23¡ãC¡À5¡ãC, less than 75%RH

DC Voltage

Range 200mV,2000mV
20V,200V,600V
Resolution 0.1mV,1mV,0.01V
0.1V,1V
Accuracy ¡À(0.8% + 2d)
Input
Impedance
10Mohm
Protection
against
overload
250V DC/AC rms
600V DC/AC rms
AC Voltage Range 200V,600V
Resolution 0.1V,1V
Accuracy ¡À(1.5%+ 5d)
Frequency 40 - 400Hz
Input
Impedance
10Mohm
Protection
Against
Overload
600V DC/AC rms
DC Current Range 20mA,200mA,10A
Resolution 0.01mA 0.1mA,
0.01A
Accuracy ¡À(1.0% + 2d)
¡À(1.5% + 5d)

Voltage
Drop

200mV
Fuse
Protection
F200mA/250V
Unfused 10A
15 second
Resistance Range 200,2000,20k,
200k,2000k,
20M
ohm
Resolution 0.1,1,0.01k,
0.1k,1k,
0.01Mohm
Accuracy ¡À(1.0% + 2d)
¡À(1.2% + 5d)
Open Circuit
Voltage
<2.8V
Protection
Against
Overload
250VDC/ACrms <30sec
Diode
Test
Test Current Approx. 1mA

Open
Circuit
Voltage

2.8 - 3.0V
Protection
Against
Overload
250VDC/ACrms <30sec
Continuity
Test
Beeper < 100 ohm

Open
Circuit
Voltage

2.8 - 3.0V
Protection
Against
Overload
250VDC/ACrms <30sec
Temperature
Internal Probe
Instrument Range 0¡æ-40¡æ
Resolution 1¡æ
Accuracy ¡À3dgt
Protection
Against
Overload
200mA/250V
fuse
Temperature
K-type probe
Instrument Range 40¡æ-400¡æ
400¡æ-1000¡æ
Resolution 1¡æ
Accuracy ¡À(0.75%+3)
¡À(2.0%+5)
Protection against overload 200mA/250V
fuse
Capacitance Measurement Range 20¦ÌF
Resolution 0.01¦ÌF
Accuracy ¡À(2.5% +5d)
Protection against overload 200mA/250V
fuse
 
BS1503